Yes, we can say the minimum amount is ≥ 10 mg/dL.
Yes, through Bluetooth connectivity.
The numbers in the System Evaluation report are reported in a different way than in the past.
Substances were tested at concentrations described by the Clinical Lab Standard Institute in document EP7-A2—Interference Testing in Clinical Chemistry; Approved Guideline, when available. Testing was conducted according to the ISO 15197:2013 standard. This is a standard used in labs. The way these concentrations are displayed will be changing, as reflected in this report. Concentrations are shown in exponential notation, in order to display values in a uniform manner across the wide range of concentrations.
To convert these values:
>Positive Exponent: Move the decimal point to the right by the number of places specified by the exponent (number after E). Acarbose Highest Concentration Tested: 6.00E+01 6.00 x 10 = 60 mg/dL
>Negative Exponent: Move the decimal point to the left by the number of places specified by the exponent (number after E). Acarbose Upper Therapeutic Concentration: 1.00E–02 1.00 / 100 = 0.01 mg/dL
>00 Exponent: Use the stated value with no adjustment. Acetaminophen Upper Therapeutic Concentration: 5.20E+00 = 5.20 mg/dL
The control solution contains a known amount of glucose that acts like blood when you apply it to the test strip. Compare the control result to the range on the label of the test strip container. If the control result is within the range, you know that the meter and test strips are working properly.
By performing a control test.